Açıklama

 

  • Specifications
  • Selection
DC Bias Current Test System
Model 6220+6632
Frequency Response 100Hz-3MHz
DC Bias Current Max. 120A
Power consumption 320W Max.
Selection
 Model 6225+6632 6240+6632 6243+6632 6223+6632 6210+6632
Frequency Response 100Hz-30MHz 100Hz-3MHz 100Hz-10MHz 100Hz-10MHz 100Hz-3MHz
Output Current Max. 20A 320A 320A 120A 60A
Power consumption 320W 640W  640W  320W  320W
units of the DC Bias  1  8  8 6  6
Output Current (unit)  20A  40A  40A 20A  10A
  • SPEC
Model 6220+6632
 Accuracy (6632) 0.000A-1.000A 1%+5mA
1.001A-5.000A 2%
5.001A-20.000A 3%
DCR Test
Constant power output
Current switch
Graphic Scanning Analysis Frequency graphic scanning analysis
Magnetic saturation curve (Isat)
Rated current curve (Irms)
  • 6632 Impedance Analyzer
Model 6632-1 6632-3 6632-5 6632-10 6632-20 6632-30 6632-50
Test

Frequency

10Hz-

1MHz

10Hz-

3MHz

10Hz-

5MHz

10Hz-

10MHz

10Hz-

20MHz

10Hz-

30MHz

10Hz-

50MHz

Equivalent

Circuit

Analysis

6632-1S 6632-3S 6632-5S 6632-10S 6632-20S 6632-30S 6632-50S
Model 6632
Frequency Range 10Hz-50MHz
Frequency Setting Continuity
Frequency Output Accuracy ±0.01%
Frequency Resolution 100mHz, 6-bit Frequency Input
Basic Accuracy ±0.08% (typical ±0.05%)
AC Drive Level Test Signal Voltage Level               10mV – 2Vrms(100Ω)

10mV – 1Vrms(25Ω)

Voltage Minimum Resolution 1mV
Accuracy      ALC ON:6% * Voltage ±2mV

ALC OFF:10% * Voltage ±2mV

Test Signal Current Level 200µA – 20mArms
Current Minimum Resolution 10µA
Accuracy      ALC ON:6% * Current ± 20µA

ALC OFF:10% * Current ± 20µA

DC Drive Level 1V (fixed)
Output Impedance 25Ω/100Ω (switchable)
DC Bias Voltage Built-in ±12V

(100Hz-50MHz)

External Voltage Bias

(±12V/±40V/±2000V)

DC Bias Current Option DC Bias Current Source

(20A/60A/120A/320A)

External Voltage Bias

(100mA/1A)

Measurement Mode Meter mode, list mode, sweep mode, and optional equivalent circuit analysis function (S model)
Measurement Parameters |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、 ESR、ε and μr
Test Time (Fastest) < 3mS
  • 6632 Measurement Parameters
Parameters Range
|Z| 0.000 mΩ to 9999.99 MΩ
R, X ± 0.000 mΩ to 9999.99 MΩ
|Y| 0.00000 μS to 999.999 kS
G, B ± 0.00000 μS to 999.999 kS
θRAD ± 0.00000 to 3.14159
θDEG ± 0.000° to 180.000°
Cs, Cp ± 0.00000 pF to 9999.99 F
Ls, Lp ± 0.00 nH to 9999.99 kH
D 0.00000 to 9999.99
Q 0.00 to 9999.99
Δ ± 0.00% to 9999.99%
Rdc 0.00 mΩ to 99.9999 MΩ
εr’ εr” 0 to 100000
μr’ μr” 0 to 100000
  • System
   Measurement Circuit Series/Parallel
Correction Open Circuit/Short Circuit/Load correction
  Cable Compensation 0 / 0.5 / 1 /2 (m)
BIN ABS、ΔABS、Δ%、OFF
   Comparator 9
 List Mode 50 groups of Multi-steps setting (Each group contains up to 15 steps)
  Bulit-in Storge 100 sets LCR setting documents, 50 groups of list mode setting
USB Host Storge LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data
  Trigger Test Auto, manual, RS-232, GPIB, Handler
   Power Supply Voltage:88-264Vac
Frequency:47-63Hz
  Power consumption 30VA
  Environment Temperature:10℃-40℃
Humidity:20-90%RH
   Dimension 337x145x525 mm (W*H*D)

/6220

336x147x340 mm (W*H*D)

/ 6632

  Weight 15 Kg (6220) 3.95 Kg (6632)
Display Color Screen, 7″ TFT (800*480)
Interface RS-232、Handler